Home Precision Environment HEPA / ULPA Filter Integrity Testing AFS 152/A Automated HEPA/ULPA Filter Scanning System
AFS 152/A Automated HEPA/ULPA Filter Scanning System product image thumbnail
AFS 152/A Automated HEPA/ULPA Filter Scanning System workflow diagram thumbnail
AFS 152/A Automated HEPA/ULPA Filter Scanning System application overview thumbnail
Topas GmbH HEPA / ULPA Filter Test Systems

AFS 152/A Automated HEPA/ULPA Filter Scanning System

Automated panel-filter scanning system for local leak detection, local-efficiency measurement, integral-efficiency determination and differential-pressure testing according to ISO 29463 and EN 1822.

Test flow
300–4,000 m³/h
Scan speed
5–100 mm/s
Specimen size
305 × 305 to 1,830 × 1,220 mm
Detection
Optical particle counter
Automated panel scan High throughput HEPA / ULPA Leak detection ISO 29463
ISO 29463-3 ISO 29463-4 ISO 29463-5 EN 1822-3 EN 1822-4 EN 1822-5

Model Overview

Model System role Primary measurements Standards / application fit
AFS 152/A Automated HEPA/ULPA panel-filter scanner Local leak detection, local and integral efficiency and differential pressure High-throughput automated testing of panel filter elements

Key Features

Automated scan traverse

A motorised probe scans the filter and frame with controlled speed and position.

Automatic concentration control

Dilution ratio and aerosol concentration can be adjusted according to filter class and flow rate.

High throughput

Automated procedures, acquisition and evaluation support production-oriented testing.

Accurate reproducible results

Controlled scan motion and particle counting reduce variability between operators.

Upgrade path

The automated traverse can be added to suitable existing AFS 152 systems.

Configurable reporting

User-defined reports document filter quality and measured test results.

Technical Specifications

Parameter Unit Value / Description
TEST PERFORMANCE
Test flow rate m³/h 300–4,000
Differential-pressure range Pa Maximum 400
Scan speed mm/s 5–100
Specimen size mm 305 × 305 × 30 to 1,830 × 1,220 × 300
SYSTEM
Test aerosols DEHS, paraffin oil and PSL
Detection Optical particle counter, 0.1–1.0 µm typical
Compressed-air supply bar 5.5–7
Power supply 230 VAC, 16 A, 50/60 Hz

Values shown are based on the standard system configuration. Confirm the final test method, specimen fixtures, instrumentation, utilities and calibration scope before order placement.

Applications

Automated panel-filter testing

High-throughput scanning of EPA, HEPA and ULPA filter elements.

Local efficiency mapping

Identification and documentation of local leakage across filter and frame.

Integral efficiency

Overall efficiency measurements according to ISO 29463-5 and EN 1822-5.

Filter-media testing

Optional workflows for filter-media efficiency under ISO 29463-3.

Similar Products